Protocol

3

Chemiluminescence detection:
HRP-tagged, alkaline phosphatase (AP)
probes or antibody conjugates

Equipment

• Clear plastic cling-wrap or clear transparent

sheet protector.

• Hoefer ULTima Pro imaging system with cooled

CCD camera.

• Pipette

Reagents (Chemiluminescence)

• ECL

(or other) detection reagent 1

• ECL (or other if required) detection reagent 2
• Membrane following hybridization

Procedure

1. Mix equal volumes of detection reagents 1 and 2.
2. Pipette the mixture over the surface of the

membrane and leave at room temperature for
1 minute.

3. Follow the wash instructions provided by the kit

manufacturer. Remove the final wash and cover the
membrane in plastic wrap.

4. Image using the Hoefer ULTima Pro chemilumines-

cence imaging system.