Ivoclar Vivadent Bluephase v.4 Instrukcja Obsługi
Strona 145
145
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 test level
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
10 V
10 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should not be used closer to any part
of the "bluephase (G2)", including cables, than the recommended separation distance
caculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 0.35 P
d = 0.35 P 80 MHz to 800 MHz
d = 0.70 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
Guidance and manufacturer's declaration - electromagnetic immunity
The “bluephase (G2)” is intended for use in the electromagnetic environment specified below. The customer or the user of the “bluephase (G2)” should
assure that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured field strength in the location in which the "bluephase (G2)" is used
exceeds the applicable RF compliance level above, the "bluephase (G2)" should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the "bluephase (G2)".
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
Rated
maximum
output power
of transmitter
W
0.01
0.1
1
10
100
150 kHz to
80 MHz
d = 0.4 P
0.04
0.13
0.40
1.3
4.0
Separation distance according to
frequency of transmitter
m
80 MHz to
800 MHz
d = 0.4 P
0.04
0.13
0.40
1.3
4.0
800 MHz to
2.5 GHz
d = 0.7 P
0.07
0.22
0.70
2.2
7.0
Recommended separation distances between portable and mobile RF communications
equipment and the “bluephase (G2)”
The “bluephase (G2)” is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the “bluephase (G2)” can help prevent electromagnetic interference by maintaining a miminum distance between portable
and mobile RF communications equipment (transmitters) and the "bluephase (G2)" as recommended below,
according to the maximum output power of the communication equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.