Maintenance and cleaning

For reasons of hygiene, it is recommended to use a
disposable protective sleeve for each patient.
Contaminated surfaces of the device and anti-glare cones
must be disinfected (e.g. FD333, FD366/Dürr Dental,
Incidin Liquid/Ecolab) and the light probe must be
sterilized before each use, unless disposable protective
sleeves are used. Make sure that no liquids or other
foreign substances enter the handpiece, charging base
and particularly the power pack during cleaning (risk of
electrical shock). Disconnect the charging base from the
power source when cleaning it.

Housing
Wipe the handpiece and handpiece holder with a cus-
tomary aldehyde-free disinfecting solution. Do not clean
with highly aggressive disinfecting solutions (e.g. solu-
tions based on orange oil or with an ethanol content of
more than 40%), solvents (e.g. acetone), or pointed
instruments, which may damage or scratch the plastic.
Clean dirty plastic parts with a soap solution.

Light probes
Clean light probe with a disinfectant and a soft cloth.
Carefully remove any possible material residue, such as
composite material, from the light probe with e.g. finger-
nail or a plastic spatula. Do not use sharp or pointed
instruments, since they may scratch the surface of the
light probe and thus reduce light transmission. Light
probes and anti-glare cones may be autoclaved (e.g.
121 °C, 2 bar, 15 minutes). Check light probe for
damage. If the light probe is held against a light source
and certain segments appear black, then the glass fibres
are broken. In this case, the light probe has to be
replaced with a new one.

Battery contacts
For good conductivity at all times, keep the battery con-
tacts free from possible contamination (e.g. composite
residue). For this purpose, clean the affected contacts
regularly in the course of the usual wipe disinfection
(after each patient).

Disposal

The apparatus must not be disposed in the
normal domestic waste. Please visit your local
Ivoclar Vivadent website for detailed informa-
tion about the disposal of the unit.
Dispose unserviceable batteries and polymer-
ization lights according to the corresponding
legal requirements in your country. Batteries
must not be incinerated.