Appendix – Ivoclar Vivadent Bluephase v.4 Instrukcja Obsługi

Strona 144

Advertising
background image

144

Appendix

Emission test

RF emissions CISPR 11

RF emissions CISPR 11

Harmonic emissions IEC 61000-3-2

Voltage fluctuations / flicker emissions
IEC 61000-3-3

Compliance

Group 1

Class B

Class A

Complies

Electromagnetic environment - guidance

The “bluephase (G2)” uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.

The “bluephase (G2)” is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for domes-
tic purposes.

P < 75W (pass without test)

P < 75W (pass without test)

Immunity test

Electrostatic
discharge (ESD)
IEC 61000-4-2

Electrical fast
transient/burst
IEC 610004-4

Surge
IEC 61000-4-5

Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11

Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8

IEC 60601 test level

± 6 kV contact
± 8 kV air

± 2 kV for power supply lines
± 1 kV for input/output lines

± 1 kV line - line
± 2 kV line - earth

<5 % U

T

(>95 % dip in U

T

) for 0.5 cycle

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

(30 % dip in U

T

) for 25 cycles

<5 % U

T

(>95 % dip in U

T

) for 5 sec

3 A/m

Compliance level

± 6 kV contact
± 8 kV air

± 2 kV for power supply lines
± 1 kV for input/output lines

± 1 kV line - line
no prot. earth

<5 % U

T

(>95 % dip in U

T

) for 0.5 cycle

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

(30 % dip in U

T

) for 25 cycles

<5 % U

T

(>95 % dip in U

T

) for 5 sec

30 A/m

Electromagnetic environment - guidance

Floors should be concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.

Mains power quality should be that of typical commercial or dental environment.

Mains power quality should be that of typical commercial or dental environment.

Mains power quality should be that of typical commercial or dental environment. If the
user of the "bluephase (G2)" requires continued operation during power mains interrup-
tions, it is recommended that the "bluephase (G2)" be powered from an uninterruptible
power supply or battery.

Power frequency magnetic fields should be at levels characteristic of a typical location in
a typical commercial or dental environment.

Guidance and manufacturer's declaration - electromagnetic emission

The “bluephase (G2)” is intended for use in the electromagnetic environment specified below. The customer or the user of the “bluephase (G2)”
should assure that it is used in such an environment.

Guidance and manufacturer's declaration - electromagnetic immunity

The “bluephase (G2)” is intended for use in the electromagnetic environment specified below. The customer or the user of the “bluephase (G2)” should
assure that it is used in such an environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.