Appendix, Emission test, Compliance – Ivoclar Vivadent Bluephase v.1 Instrukcja Obsługi

Strona 142: Electromagnetic environment - guidance, Compliance level

Advertising
background image

142

Appendix

Emission test

RF emissions CISPR 11

RF emissions CISPR 11

Harmonic emissions IEC 61000-3-2

Voltage fluctuations / flicker emissions
IEC 61000-3-3

Compliance

Group 1

Class B

Class A

Complies

Electromagnetic environment - guidance

The bluephase style uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.

The bluephase style is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.

P < 75W (pass without test)

P < 75W (pass without test)

Guidance and manufacturer‘s declaration – electromagnetic emission

The bluephase style is intended for use in the electromagnetic environment specified below. The customer or the user of the bluephase style should
assure that it is used in such an environment.

Guidance and manufacturer‘s declaration – electromagnetic immunity

The bluephase style is intended for use in the electromagnetic environment specified below. The customer or the user of the bluephase style should
assure that it is used in such an environment.

Immunity test

Electrostatic
discharge (ESD)
IEC 61000-4-2

Electrical fast
transient/burst
IEC 61000-4-4

Surge
IEC 61000-4-5

Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11

Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8

IEC 60601 test level

± 6 kV contact
± 8 kV air

± 2 kV for power supply lines
± 1 kV for input/output lines

± 1 kV line - line
± 2 kV line - earth

<5 % U

T

(>95 % dip in U

T

) for 0.5 cycle

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

(30 % dip in U

T

) for 25 cycles

<5 % U

T

(>95 % dip in U

T

) for 5 sec

3 A/m

Compliance level

± 6 kV contact
± 8 kV air

± 2 kV for power supply lines
± 1 kV for input/output lines

± 1 kV line - line
no prot. earth

<5 % U

T

(>95 % dip in U

T

) for 0.5 cycle

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

(30 % dip in U

T

) for 25 cycles

<5 % U

T

(>95 % dip in U

T

) for 5 sec

30 A/m

Electromagnetic environment - guidance

Floors should be concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.

Mains power quality should be that of typical commercial or dental environment.

Mains power quality should be that of typical commercial or dental environment.

Mains power quality should be that of typical commercial or dental environment. If the
user of the bluephase style requires continued operation during power mains
interruptions, it is recommended that the bluephase style be powered from an
uninterruptible power supply or battery.

Power frequency magnetic fields should be at levels characteristic of a typical location in
a typical commercial or dental environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.