Programming the TIME key

1

Press the TIME key.

2

Enter a value using the numeric keypad. The LED will
display the new value.

3

Press the ENTER key. After programming, pressing
the TIME key retrieves this setting. Change the setting
by entering a new value and pressing ENTER.

Stopping and restarting the
crosslinker

• Press the red STOP/RESET key to stop the

exposure. The LED will display the remaining
exposure level or time.

• Press the green START key to restart the expo-

sure. The program will continue from where it
was stopped.

• To enter a different value once the program has

started, press STOP/RESET, enter the new value,
then press ENTER. Press the START key to begin
the new exposure value.

• To revert to the last used setting after the expo-

sure is interrupted, press the STOP/RESET key a
second time. Press START to start exposure from
the previous setting.

• If the door is accidentally opened during expo-

sure, the UV lamps instantly turn off and the
remaining exposure level or time is displayed on
the LED display. To restart, close the door and
press the START key.