Operating instructions – Hoefer UVC500 Instrukcja Obsługi
Strona 14
Operating instructions
Set up and membrane placement
1
Place the unit on a stable and level surface.
2
Plug the power cord into the receptacle at the back of
the unit and into a properly grounded electrical outlet
of the proper power rating.
3
Flip the power mains switch located under the keypad
to on (–).
Note: The instrument settings are always the values
last programmed, and are indicated by the LED
display and a lit red indicator lamp.
4
Open the door by pulling the door handle above the
window toward you.
Note: Do not obstruct or damage the photofeed-
back sensor, which is mounted on a small metal tab
protruding from the lower right side.
5
Moisten (do not wet completely) a sheet of blotting
paper with transfer buffer and place it in the center of
the compartment. Place the membrane on the paper
with the nucleic acid side facing upward.
6
Set the energy exposure (energy or time) as described
in the next section.
•
p5
Note: To protect the cross-
linker from corrosive buffers,
lay a piece of plastic wrap
on the compartment bottom
before placing the blotting
paper and sample.