Hoefer UVC5000 Instrukcja Obsługi

Strona 14

Advertising
background image

The UV Crosslinker can be operated on
the following settings:

Preset UV energy exposure setting

1

Push the PRESET and then the ENERGY button on
the tactile touch pad. The red light at each position
should now be lit and the preprogrammed UV energy
exposure setting of 120,000 μJ/cm² is displayed in
the LED as 1200.

2

Push START on the touch pad. After a slight delay to
energize the UV tubes, the LED will begin to count
down. The unit will automatically stop at the end of
the exposure cycle and will beep five times.
Exposure is now complete.

Optional: Though the preset UV energy exposure
setting is factory set at 120,000 μJ/cm², it is possible
to change this if necessary, as follows:
• Push and hold the PRESET button on the touch

pad until you hear an audible signal, then push
the ENERGY button on the touch pad.

• Set your new preset UV energy exposure setting by

inputting the numbers on the touch pad. The new
setting will appear on the LED display.

• Push ENTER on the touch pad. The new setting

is now established.

p5

Note: The LED displays 1200.
This must be multiplied by 100
to obtain the UV exposure value.

Note: The LED display must
be multiplied by 100 to
obtain the actual exposure
setting.

Advertising