Calibration, Product specifications – Braun ThermoScan IRT 3020 CO Instrukcja Obsługi
Strona 7
7
Calibration
The thermometer is initially calibrated at the
time of manufacture. If this thermometer is used
according to the use instructions, periodic
readjustment is not required. If at any time you
question the accuracy of temperature measure-
ments, please contact Braun Infoline (see page 2).
Manufacturing date is given by the LOT number
located in the battery compartment. The first
number after LOT represents the last digit of the
year of manufacture.
The last two digits give the week of manufacture.
An example: LOT 503 – this product has been
manufactured in the week 03 of the year 2005.
Product specifications
Displayed temperature
range:
34 °C – 42.2 °C
(93.2 °F – 108 °F)
Operating ambient
temperature range:
10 °C – 40 °C
(50 °F – 104 °F)
Display resolution:
0.1 °C or °F
Accuracy for displayed
temperature range:
± 0.2 °C (35.5–42 °C)
(95.9–107.6 °F)
± 0.3 °C (outside this
temperature range)
Clinical repeatability:
± 0.25 °C (± 0.45 °F)
Humidity:
≤
95% non-condensing
Battery life:
2 years /
1000 measurements
If device is not used within specified temperature
and humidity ranges the technical accuracy can
not be ensured.
Type BF equipment
Subject to change without notice.
This appliance conforms to the following
standards:
DIN EN 60601-1: 3/96 «Medical electrical
equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers» –
Part 5: Performance of infrared ear thermometers
(with maximum device)
This product conforms to the provisions
of the EC directive 93/42/EEC
(Medical Device Directive).
MEDICAL ELECTRICAL EQUIPMENT needs
special precautions regarding EMC. For detailed
description of EMC requirements please contact
the Braun Infoline (see page 2).
Portable and mobile RF communications
equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Please do not dispose of the product in the
household waste at the end of its useful life.
Disposal can take place at a Braun Service
Centre or at appropriate collection points
provided in your country.
0297
I6026223_IRT-3020_Nordic_2 Seite 7 Mittwoch, 26. Juli 2006 11:32 11