Microlife BP 3BT0-H Instrukcja Obsługi

Strona 11

Advertising
background image

5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measure-
ment for any reason (e.g. the patient feels unwell), press
the quick release valve at the pump ball. The device then
immediately lowers the cuff-pressure automatically.

5.7. Memory – displaying the last measurement
The blood-pressure monitor automatically stores the last
measurement value. By pressing the MEMORY button for
at least 3 seconds, the last measurement will be
displayed.

6.

Error messages/malfunctions

If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(example: error no. 1)

Error No.

Possible cause(s)

ERR 1

The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
This situation can occur, for example, if the tube becomes unplugged after the systolic blood-
pressure has been measured. Further possible cause: No pulse has been detected

ERR 2

The pressure dropped below 20 mmHg without a measurement taking place. The Re-
lease valve was pressed during the measurement.

ERR 5

The measured readings indicated an unacceptable difference between systolic and dias-
tolic pressures. Take another reading following directions carefully. Contact you doctor if
you continue to get unusual readings.

HI

The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes and repeat the measurement.*

LO

The pulse is too low (less than 40 beats per minute). Repeat the measurement.*

* Please consult your doctor, if this or any other problem occurs repeatedly.

Other possible malfunctions and their elimination
If problems occur when using the device, the following points should be checked and if necessary, the
corresponding measures are to be taken:

8

MR

3 sec.

Malfunction

Remedy

The display remains empty when the instrument is
switched on although the batteries are in place.

1. Check whether the batteries are installed with the

correct polarity and correct if necessary.

2. If the display is unusual, remove the batteries and

then exchange them for new.

Advertising